Description
The HV Diagnostics HVA60 provides very-low-frequency AC and DC high-voltage testing for cable insulation assessment.
Selectable VLF sinewave and DC output modes support cable withstand testing and other compatible insulation test applications. Automatic and manual test sequences, digital voltage/current metering and stored results support documented field testing. Tan delta and partial-discharge diagnostics require compatible external equipment and are not implied by the base model. Choose test voltage, waveform and accessories for the cable construction and applicable test procedure.
Key Specifications
- Maximum sinewave output: 44 kV RMS / 62 kV peak
- DC output: ±60 kV
- VLF frequency: 0.01–0.1 Hz
- Control: Automatic and manual test sequences
- Duty cycle: Continuous operation
- Capacitive load capability: 1 µF at 44 kV RMS and 0.1 Hz
- Input supply: 110–240 V AC, 50/60 Hz
Request a rental or sales quote to confirm availability, configuration and included accessories.