Description
The 3-037-R002 moisture measuring device was designed, from the ground up, to be used on arced SF6. Unlike conventional multi-gas hygrometers, the 3-037 is specifically calibrated for the unique characteristics (i.e. densities, molecular weights, and acidity) of faulted SF6. Not only is the device consistently more accurate and reliable, but the typical long-term drift (and subsequent need for factory recalibration) associated with non gas-specific devices is all but eliminated.
The exclusive DILO sensor measures SF6 at atmospheric pressure which removes pressure and temperature variables from moisture readings. Additionally, the electronic flow meter not only reduces the sample to a trickle, but the orientation of the device is no longer relevant – making the 3-037 far more suitable for on-site testing. All readings are displayed in parts per million by volume (PPMv).
STANDARD VERSION
Moisture measuring device with digital display
2 precision regulation valves
Powered by AC or NiMH batteries (internal charger)
2 meter long connecting hose with couplings DN 8 and DN 20
Rugged case with handle
2 meter long power cable (for AC use)
Aluminum carrying case
2 operating instructions in: ( ) German ( ) English ( ) French
TECHNICAL DATA
measuring range: 41p PPMv – 10,000 PPMv
measuring accuracy: ± 2 °C (±4 °F)
max. input pressure: pe 10 bar (145 PSIG)
operating temperature: 0 – 40 °C (32 °F – +104 °F)
operating voltage: 100 – 240 V 50/60 Hz
flow rate 1 liter of SF6/min